Process mode is the workspace for reviewing, reprocessing, saving, and exporting previously recorded Vimotis data. Use this mode when an acquisition file already exists and you need to inspect signals, adjust processing settings, add markers, export data, or compare channels across files.
Process mode supports:
- loading recorded
.h5or.hdf5files, - reviewing wave-speed, calibrated-load, analog, correlation, and foot-contact views,
- updating processing parameters and calibration data,
- adding marker points to selected channels,
- saving reprocessed results back into the loaded HDF5 file,
- exporting CSV outputs and figures,
- running batch reprocess or export workflows, and
- comparing channels across files in the Evaluation window.
Always post-process tensiometer data before detailed analysis or exporting. The processing algorithms which allow for live data display are less robust than those used to post-process the data.
Menus and Dialogs
When Process mode is active, these unique process mode menu actions are available:
- File > Load: load a Vimotis HDF5 file.
- File > Save: save reprocessed results into the loaded HDF5 file.
- File > Export: open the CSV export dialog.
- File > Batch: open the Batch Process workflow.
- Edit > Calibration: open calibration settings for Process mode.
- Edit > Parameters: open normal processing parameters.
- Evaluate > Open: open the Evaluation window.
Loading Data and Sources
The application loads HDF5 files with .h5 or .hdf5 extensions. If a file contains both live and reprocessed results, the user will be prompted with a pop-up to choose whether they wish to load Live data (as collected and viewed in real-time) or Reprocessed (post-processed and saved data).
The displayed results use the calibration saved with that result. For a linked archive, Reprocess All initially uses the Live calibration when Live results are open, or the archive default when Reprocessed results are open. If these differ from the saved result's calibration, a message identifies the calibration that reprocessing will use.
Additionally, if a file is already loaded in the workspace, and the user chooses to load Live data, a dialog will pop up with options Keep Current or Reset Defaults.
Keep Currentallows the user to keep the processing parameters they set/loaded from the previous file in the workspace. Note that to apply these parameters to the new file, the user will need to runReprocess All.Reset Defaultsresets the workspace parameters to the default reprocessing parameters.
Save Data
Reprocessed data is not saved by default to prevent loss of desired data when viewing or experimenting with different processing parameters. To save reprocessed data, choose File > Save. Reprocessed data is stored in the same file as the Live data. Only one set of Reprocessed data can be saved per file; any future saved reprocessing will overwrite old Reprocessed data. Once Reprocessed data has been saved, it may be chosen for viewing/reprocessing when loading that file in the future.
An asterisk after the filename indicates unsaved run changes. Choose File > Save to retain them; the filename then shows (saved). The indicator does not track an unapplied calibration draft.
Workspace and Plot Controls
The main Process workspace includes a left-side channel selection area and plot tabs for reviewing the loaded file.
- Switch between
Wave SpeedandCalibrated Loadusing the dropdown in the upper-right of the main plot tab. - Plot items can be hidden/revealed by clicking on the legend entry to toggle visibility.
- The x- and y-axis zoom is controlled by clicking and dragging in the direction and over the region of interest. Double-click to reset the axes.
Channel Select

The left-side channel control area is used to choose which channels are visible and which channel is active for marker editing.
- Channel check boxes determine which wave-speed and analog signals are visible.
- The Add Markers controls choose the active wave-speed or analog channel for marker editing.
- Sets the low-pass filter cutoff used for wave-speed smoothing. The filter is a zero-phase second order Butterworth filter with a maximum cutoff of 50Hz (Nyquist).
- The
Reprocess Allcontrol recalculates selected wave-speed outputs with the current processing settings. - Analog data acquired on auxiliary AI ports are displayed on separate axes below the load plot. Optional low/high/bandpass filtering (2nd order Butterworth) can be applied to these data on a per-channel basis.
- For relevant datasets, foot contacts can be applied. These are estimated using alternative processing of the raw tensiometer data and should only be used as estimates.
Correlation Image
The core data processing algorithm computes a wave speed based on a changing time delay (Δt). The time delay represents wave travel time between the two sensor contacts and is found using cross-correlation based algorithm. This time delay (shown in red), overlayed on a heatmap representation of the two sensor contact signal correlations can be viewed in the Correlation tab.

The x-axis indicates the tap index, or the index of the excited wave. The tap rate is 100 Hz, so the number of indices should be 100x the duration of the collection in seconds. The y-axis indicates the full length of possible time delays (Δt) values for a given tap n. The heatmap indicates the correlation strength at each possible time delay, with the final time delay following the peak values in the correlation image. Since the waves are oscillatory, multiple peaks can be observed, and the processing algorithm determines the appropriate peak trace from which to compute wave speed and load. Gaps indicate data that were excluded by the quality gate settings. For tap indices without any high correlation coefficients (yellow in the plot), shape similarity between the upstream and downstream signals is low, so confidence in the Δt value is low.
Foot Contact
For gait tasks, foot contacts may be optionally estimated. To enable, check the foot contacts box in the left channel bar. The channel source for the foot contacts may be selected from the dropdown, or each channel may be assigned a foot contact source if Per-Channel Sources is selected.

- Choose which channel is shown for foot contact estimation from the dropdown.
The foot contact detection uses features in an accelerometer signal to estimate ground impact. Depending on the task and sensor location, the parameters to isolate foot contact features may require optimization. The available parameters are:
- Band Low/High [Hz]: Bandpass filter cutoffs.
- Prominence: Amplitude of feature peak relative to local minima.
- Refractory [s]: Minimum permitted time between foot contact events. For example, during high-cadence running, the time between events will be much shorter than walking, so a shorter refractory should be used.
- Foot contact event detection may be edited directly by single-clicking on peaks to add or remove them. Recomputing foot contacts after changing parameters will remove manually edited events.
- Switch between point interaction and zoom mode using the respective buttons.
Data Markers
Markers may be added per-channel to denote timing or specific features of interest. These are exported in a column vector time-synced with the rest of the data for external processing. Note that in marker mode, the standard x/y zoom interaction is toggled with the associated button.

Markers are saved with reprocessed results and included in CSV exports when the Markers export option is selected.
Process Parameters Dialog
The Process Parameters window supports per-sensor editing of processing settings.

| Setting | Description |
|---|---|
| Apply changes to all sensors | Applies the visible parameter changes across all tensiometer channels. Uncheck to change individual channel settings independently. |
| Tensiometer tabs | Select the sensor whose processing settings are being reviewed or edited. These are only enabled when Apply changes to all sensors is disabled. |
| WS LP [Hz] | Low-pass cutoff for wave-speed smoothing. Reprocess after changing it. |
| Quality Gate | The Quality Gate is used to mask data in which there is low confidence based on two parameters. Min Corr is the minimum allowable cross correlation coefficient found when computing the wave travel time (Δt) between the two sensor contacts; it is a measure of distortion, or lack thereof, in the signal as it is transmitted from contact 1 to contact 2. Too much distortion between the signals lowers confidence in the computed Δt (see Correlation Image and Section 7.4 of the Product Manual). Min Amp Ratio is the minimum ratio between the maximum amplitude of the signal measured at contact 2 divided by contact 1; it represents how well the signal is being transmitted through the tendon. |
Use Reprocess All in the main app window after changing parameters when the loaded data should be recalculated with the updated settings.
Create Calibration from Recorded Data
In Process, open Edit > Calibration, then choose Create from recorded data… to create a calibration using recorded tensiometer data and external joint torque.
Prepare and Map Sources
- Add tensiometer
.h5recordings containing raw accelerometer data. - Add torque CSVs in UTF-8 format. The first column must be
time_sortime_ms, followed by uniquely named torque columns ending in_Nm, for exampletime_s,left_ankle_Nm,right_ankle_Nm. Times must be increasing and unique; torque values must be in N·m. - Enter participant bodyweight, then select each sensor tab and complete its source and calibration settings.

| Setting | Description |
|---|---|
| Bodyweight [N] | Weight of the participant in newtons. Applies to all included sensors. |
| Include sensor | Check to include the sensor in the calibration. Uncheck sensors that should be excluded. |
| Raw channel | Select the tensiometer recording and channel for this sensor. |
| Torque file | Select the matching external torque CSV. |
| Torque column | Select the named joint-torque column for this sensor. |
| Moment arm [cm] | Approximate tendon moment arm, assumed fixed during the calibration. |
| Contribution [%] | Percentage of joint torque attributed to the measured muscle-tendon group. For example, enter 50 for 50%. |
| Sign | Choose As recorded to keep the torque sign or Invert sign to reverse it. |
Align and Fit
Alignment and fit settings are configured separately for each sensor.
- Select the sensor tab. Choose Auto-align or adjust Offset manually, then check that the corresponding movement features line up in the Alignment plot.
- Select the fit interval and review the fit criteria.
- Choose Fit calibration, then review the Calibration fit plot and fit results. Fit again after changing sources, alignment, interval, or fit settings.

| Setting | Description |
|---|---|
| Offset [s] | Time offset between the torque and tensiometer recordings. Positive values move torque later. Edit the value to adjust alignment manually. |
| Auto-align | Sets an alignment offset automatically. Review the plotted traces before fitting. |
| Search range | Expands the Global ± and Anchor ± range settings. |
| Global ± [s] | Alignment search range when no manual offset has been entered. |
| Anchor ± [s] | Alignment search range around a manually entered offset. Increase if the appropriate alignment lies outside the current range. |
| Clear | Resets the offset to zero and clears the manually selected alignment. |
The Fit area and Fit criteria contain the calibration settings:
| Setting | Description |
|---|---|
| Interval [s] | Start and end times of the portion of the recording to use for calibration. Select the intended calibration task and load range. |
| Fit criteria | Expands the fit acceptance settings listed below. |
| MAD threshold | Controls how strictly outliers are excluded from the calibration fit. Lower values use a stricter threshold. |
| Minimum R² | Minimum fit correlation strength required for acceptance. |
| Minimum samples | Minimum number of valid samples remaining after exclusions for an accepted fit. |
| Minimum load [BW] | Minimum tendon load to include, as a multiple of participant bodyweight. |

Add a Separate Zero-Load Recording
Adding a separate zero-load recording is recommended but optional.
- After fitting, expand Zero-load capture and choose Load zero-load…. Select a separate
.h5recording collected with the tendon unloaded. - Select the sensor and unloaded interval. Review the trace, mean wave speed, and valid tap count, then choose Use capture.
- After choosing Use capture, the zero-load value is enabled in the creator. If the fitted zero-load value is unavailable, a valid capture must be selected and enabled before saving.
- If you did not capture a separate zero-load recording, an unloaded phase of gait data can sometimes be used instead. Load the gait recording and select an unloaded interval.

| Setting | Description |
|---|---|
| Source sensor | Select the sensor channel in the unloaded recording. Use the same sensor setup as the calibration recording. |
| Unloaded segment [s] | Start and end times of the unloaded interval on this recording. |
| Use capture | Accepts the selected unloaded interval and returns to the creator. |
Save and Use the Archive
Choose Save archive… and enter an unused filename. Existing files cannot be overwritten. Load the saved archive in Process or Batch to use the calibration.
Calibration Workspace
Open Edit > Calibration to review or edit a run's calibration. A linked calibration archive is normally found beside the run file; use the calibration workspace to load an archive from another location. Check the sensor mapping when selecting a different archive.
Before updating an archive, review the compatibility and backup guidance if you need to use it with an older app version.

Select Review for a sensor to open its workspace.

Edit a sensor, choose Recompute Calibration, review the fit and any zero-load requirement, then choose Save to Draft. Repeat for other sensors as needed. In the summary, choose Save and Apply to save the complete set as Modified and make it the archive default. This overwrites the previous Modified set; Live or Original remains unchanged. To change the archive default without editing sensors, select the desired set in the summary and choose Save and Apply. In Process, applying a calibration automatically reprocesses the open run in memory; use File > Save separately to retain its Reprocessed Results. A saved archive remains changed if the subsequent reprocessing fails.
Seed from Live restores the selected sensor's Live calibration to its workspace. Choose Save to Draft and then Save and Apply to retain that change in Modified.
Reprocessing uses the applicable linked archive or an offered run calibration snapshot. If an archive does not cover the active sensors, review the warning and offered snapshot before continuing. If usable calibration is missing, the application may offer a nominal estimate or cancellation. Live data and its calibration snapshot remain unchanged.
The Calibration fit and Measurements tabs show the fit and, when available, reference load and measured wave speed squared over time. For a hardware-reference archive with suitable data, Auto-align estimates the analog-input sync delay. Review the estimate, recompute the fit, and apply it. Clear resets the sync delay. The recorded-data calibration creator uses a separate alignment offset.

CSV Export Dialog
File > Export opens the CSV export dialog.

| Option | Description |
|---|---|
| Wave speed | Exports wave-speed data from the selected source. |
| Calibrated load | Exports calibrated-load data when calibration data are available. |
| Analog channels | Includes checked analog channels on the wave-speed time base. |
| Foot contact | Exports foot-contact data when available or generated. |
| Markers | Includes saved marker locations in the export. |
| Full-rate analog channels | Writes full-rate analog data to a separate CSV. |
Select Standard export, Export time-normalized strides (separate CSV), or both. Standard export is selected by default. Full-rate analog data are written to a separate standard CSV.
Time-normalized export requires foot contacts enabled in the workspace and at least one of wave speed, calibrated load, analog channels, or markers. The _timenorm.csv contains each eligible stride as 101 points from 0% through 100%, with separate percentage columns for groups with different contact boundaries. It includes all eligible strides; Evaluate's display-time outlier exclusions do not filter the export.
The application warns before overwriting existing CSV files.
Batch Process Dialog
File > Batch opens a guided workflow that can run in two modes: Reprocess Files and Export Existing CSV Only. Export Existing CSV Only is intended for use when only the export function is required. The workflow is organized into four steps:
Select Files: choose a source folder and move files into the
Files To ReprocessorFiles To Exportlist.Select files and use the arrow buttons to move selected files or all files between lists. Green filenames indicate saved Reprocessed data. In reprocess mode, existing Reprocessed data in selected files will be overwritten.

Figure 14: Batch file selection with an example folder and filenames. Confirm Parameters: review or seed processing parameters when reprocessing. This step is skipped in
Export Existing CSV Onlymode.Option Description Use Defaults Reprocesses using standard settings. Seed From File Applies the processing parameters from the selected file to all files. May choose between Live and Reprocessed if available. Seed From Workspace Applies the processing parameters from the workspace. The intended use case is: load an individual file to the workspace. Modify processing parameters until desired results are achieved. Run the batch wizard and apply workspace parameters to batch.
Choose calibration independently from processing-parameter seeding:
| Calibration source | Behavior |
|---|---|
| Use one calibration archive | Uses the selected archive's current default for the batch. Review archive and sensor mappings first. |
| Use run file snapshots | Uses each run's saved calibration: Reprocessed when available, otherwise Live. Supports different saved calibrations across runs; does not refit an archive. |
In archive mode, locate the linked archive or browse for one. View Parameters opens the calibration workspace to review or edit the shared calibration. If selected files reference different calibration sets, browse for one archive that covers the batch. Resolve missing archives or incompatible mappings before running. If required calibration is missing, reprocessing may offer a nominal estimate or cancellation.

In shared archive mode, save sensor edits to the draft and choose Save and Apply in the summary before running Batch. The Process calibration review illustrates the shared workspace. Batch does not reprocess its selected files until it runs. Auto-align is available for compatible hardware-reference archives, but not when using run snapshots or imported external torque calibrations.
Save and Export: choose the batch mode, data source, CSV outputs, summary plots, and foot-contact options.
These options mirror those in the Export dialog.
Time-normalized export requires an eligible selected data type and foot contacts in every selected file, or enabled (Re)estimate foot contacts. In export-only mode, if estimation is unavailable, every selected file must already have contacts. Review any skipped or unusable data reported by the batch.
Additionally, choose whether to compute foot contacts for the batch files. As with the workspace Foot Contact options, these can be sourced from a single channel for all channels (e.g. all sensors/signals associated with the right leg), or applied independently to each channel (e.g. bilateral measurements).
An optional summary plot may be selected to show all processed data after the batch runs.

Figure 16: Batch Process CSV output and foot-contact options.
In Export Existing CSV Only, choose the existing result source:
| Export source | Behavior |
|---|---|
| Prefer Reprocessed | Uses saved Reprocessed data if present, otherwise Live. Review reports files that fall back to Live. |
| Reprocessed Only | Requires saved Reprocessed data; files without it fail rather than falling back. |
| Live Only | Uses Live data. |
Export-only does not reprocess or write the HDF5 files. Reprocess Files can save new results and export them in the same run.
Options include downsampled analog in the main CSV and full-rate analog in a separate CSV.
- Review and Run: confirm the selected files, parameter/calibration sources, and export options before starting.
Review the summary before running the batch. In Reprocess Files mode, previously saved Reprocessed data will be overwritten; choose Run Reprocess to start. In Export Existing CSV Only mode, choose Export CSV. The progress bar at the bottom of the wizard shows the batch progress.

Normal Output
Saving in Process writes Reprocessed Results, processing parameters, and any generated foot contacts or markers into the loaded HDF5 file. The application prompts before overwriting existing Reprocessed Results.
CSV filenames include the loaded file stem and the selected data-source token:
- Main CSV:
{stem}_{source_token}_postprocess.csv - Full-rate analog CSV:
{stem}_{source_token}_postprocess_full_ai.csv - Time-normalized stride CSV: a separate file ending in
_timenorm.csv
The source token identifies the actual exported source (Live or Reprocessed).
Evaluation Window
The Evaluation window can be accessed via the Evaluate > Open menu. This tool is intended to allow for quick comparisons of data across one or more files and should not be a substitute for a more thorough data processing pipeline.

Add Files...loads one or more HDF5 files into the comparison set.Remove Selectedremoves highlighted files from the comparison set.Clearremoves all loaded files.Displaydropdown toggles between wave speed and load.Export Figureopens the Export Figure dialog.
File/channel tree
- Check/uncheck boxes to add/remove channels on the plot. Double-click on a channel name to assign an alias to the channel. This alias does not change the underlying data but can be useful to clarify the plot legend.
Analog channels are plotted on a separate axis below the wave speed / load axis.
Selected Channels
- Shows the active comparison channels. Drag and drop items in this list to reorder the plot entries and legend.
Stride View
- If foot contacts have been computed for the loaded files, checking the box will time-normalize each stride to 0-100% gait cycle.
- Strides may be plotted in aggregate (all strides not eliminated by the outlier threshold) or as an ensemble average (+/-1SD).
Exclude Outlier Strides removes strides when more than the selected percentage of points differ by over 3 SD from the other strides. Adjust the percentage threshold to control how many outlier points cause exclusion.
Exclude Strides with Outlier Peaks excludes a stride whose peak differs by more than 3 SD from the other stride peaks, regardless of peak timing. With both options enabled, strides excluded by either are omitted from the plot and ensemble statistics.
Align by event is an alternative to foot-contact based alignment for non-repetitive tasks (e.g. jumping, pitching). Choose
Configureto launch the Align by Event wizard.Configure event alignment
- Alignment: choose
One PointorTwo Points.One Pointidentifies a single point in each channel and translates all traces such that the selected point in each channel is aligned.Two Pointsboth translates and linearly warps each channel so that both selected points are aligned across channels. - Point Type: options are
Peak,Trough, orManual.PeakandTroughwill identify the nearest feature of matching type when selecting.Manualwill place the point at the location clicked. - When plotting multiple channels from the same file, you may choose to apply the alignment point(s) across those channels or set them independently.
- Alignment: choose

Figure 19: Align by Event wizard, configure event alignment. Choose points
For each channel, select the desired alignment point(s), using
Nextto proceed through each plot. To adjust a selected point, use the upper-left dropdown to choose which point you want to adjust, then reselect the correct location. Once all points have been selected for all channels, chooseFinish

Figure 20: Align by Event wizard, choose points - Once the pipeline is complete, the checkbox will be enabled to toggle the event alignment.
Export Evaluation Figure Dialog

The Export Evaluation Figure dialog is used to configure plot output before saving a comparison figure.
- Output type: Selects
BitmaporVector (SVG). - Plot title and axis labels: Sets the title and axis text used in the exported figure. These are applied only in the export preview and will not impact the main Evaluate window plot.
- Axis ticks: Set location of axis ticks. To auto-fill an evenly spaced range, use the syntax [low:step:high].
- Font sizes: Independently control font size for title, labels, and legend.
- Legend placement: Sets where the legend appears in the exported figure.
- Export theme: Selects the visual color theme used for output,
DarkorLight Size preset: Options for exported figure size, selects
Half-Slide,Full-Slide, orManual.Half-SlideandFull-Slideare based on standard PowerPoint slide dimensions. IfManualis chosen, additional options are exposed:- Units: Selects pixels, millimeters, or inches.
- Dimensions: Width and height of exported figure in specified units.
- DPI: Controls bitmap export resolution (if Output type is
Bitmap)
- Preview: Shows a live preview of the configured figure (resolution will be lower than exported image).
Vector export saves SVG. Bitmap export supports PNG and JPEG.